ru-RUuk-UAen-GB

Avantgarde.Line | Series FD with forced convection

Full description

  • Temperature range: ambient temperature plus 10 °C to 300 °C

  • High temperature accuracy thanks to APT.line™ technology

  • Forced convection

  • Controller with LCD display

  • Electromechanical control of the exhaust air flap

  • 2 chrome-plated racks

  • Units up to 115 L are stackable

  • Class 2 independent temperature safety device (DIN 12880) with visual temperature alarm

  • Ergonomic handle design

  • USB port for recording data



BINDER heating chambers with forced convection for quality and reliability
BINDER heating chambers boast impressively high levels of quality and reliability. Ideal for drying, sterilization, and heating, they can be put to use within industry, science and research, and quality assurance applications. Alongside models with natural convection, BINDER also offers heating chambers with forced convection. These heating chambers – in the Classic.Line FD series – are a popular choice when extremely fast drying and sterilization are required.
BINDER's heating chambers with forced convection operate within a temperature range of between 5 °C and 300 °C, allowing them to be adapted to suit various requirements. The patented APT.line™ preheating chamber technology guarantees homogeneous temperature distribution in the interior, so you can be sure that sample conditions are consistent across the entire usable space, regardless of the amount or size of the sample. Low heat dissipation means that the heating chambers are highly energy-efficient despite performance being increased.

BINDER heating chambers for efficiency, high performance, and ease of use
BINDER's FD series brings you maximum quality and precision when the heating chambers are put to use in practice, guaranteed by high production standards in line with DIN 12880 along with premium material and finish quality. What's more, the chamber can be cleaned with ease thanks to the stainless steel finish of the interior as well as the lack of built-in fixtures there. To facilitate loading and unloading, the heating chambers are fitted with racks that come with anti-tilt protection.
With BINDER heating chambers with forced convection, not only are you getting the benefits of their high performance, functionality, and process reliability, you can also enjoy these units' excellent usability, as they are intuitive and easy to work with.
Product specification
Model FD 56 FD 115 FD 260 FD 720
Measures
Interior volume [L] 60 116 259 741
Net weight of the unit (empty) [kg] 39 50 85 166
Permitted load [kg] 70 150 270 315
Load per rack [kg] 30 30 40 45
Internal Dimensions
Width [mm] 400 550 650 1000
Height [mm] 440 550 780 1300
Depth [mm] 345 385 515 570
Doors
Unit doors 1 1 1 2
Outer dimensions
Width net [mm] 560 710 810 1165
Height net [mm] 625 735 965 1590
Depth net [mm] 565 605 760 870
Wall clearance back [mm] 160 160 160 160
Wall clearance sidewise [mm] 100 100 100 100
Environment-specific data
Energy consumption at 150 °C [Wh/h] 290 340 410 800
Sound-pressure level [dB(A)] 43 43 43 43
Fixtures
Number of shelves (std./max.) 2/4 2/5 2/8 2/16
Performance Data Temperature
Temperature range [°C] above ambient temperature 10 10 10 12
Temperature range to [°C] 300 300 300 300
Temperature variation at 150 °C [± K] 1.7 1.7 1.9 2.5
Temperature fluctuation at 150 °C [± K] 0.3 0.3 0.4 0.6
Heating-up time to 150 °C [min] 15 18 19 25
Recovery time after 30 seconds door open at 150 °C [min] 4 4 5 6
Electrical data
Rated Voltage [V] 230 230 230 400
Power frequency [Hz] 50/60 50/60 50/60 50/60
Nominal power [kW] 1 1 2.3 4.5
Unit fuse [A] 6.3 6.3 12.5 3 x 16
Phase (Nominal voltage) [ph] 1~ 1~ 1~ 3~
All technical data is specified for unloaded units with standard equipment at an ambient temperature of +22 °C ±3 °C and a power supply voltage fluctuation of ±10%. The temperature data is determined in accordance to BINDER factory standard following DIN 12880, observing the recommended wall clearances of 10 % of the height, width, and depth of the inner chamber. All indications are average values, typical for units produced in series. We reserve the right to change technical specifications at any time.